Xtend Magnet System


Magnet weight 4600 kg
Open bore diameter 70 cm
Maximum FOV 55 cm
Typical homogeneity at 55 x 55 x 50 cm ≤ 5 ppm
Typical homogeneity at 50 x 50 x 45 cm ≤ 1.8 ppm
HeliumSave technology Yes (Zero boil-off)
Cryogen boil-off rate

0.0 l/hr*

 

*under regular scanning conditions

Xtend Gradient


3.0T Omega HP
Max. amplitude for each axis 45 mT/m
Max. slew rate for each axis 200 T/m/s

Resolution Parameters


3.0T Omega HP
Max scan matrix 1024 (2048 optional)
Highest in-plane resolution 5 μm
Max. number of slices 1024

RF Transmit


Parallel RF transmission Yes, MultiTransmit
RF control Patient-adaptive. Fully flexible for each source, real time feedback loop

dStream RF Receive


Number of independent receive channels Channel Independent
Location of analog-to-digital converter (ADC) Inside the coil
Signal chain from coil to reconstructor

 

Signal chain from coil electronics to connector Digital
Signal chain from connector to magnet Digital
Signal chain from magnet to reconstructor Digital

Patient Environment


Patient aperture 70 cm
Flare on both ends Yes
Tunnel diameter at both ends 95 cm
Maximum weight capacity* 250 kg (550 lbs)
Patient transport system (optional)

FlexTrak
FlexTrak Mammo

Wireless patient physiological synchronization Yes
Various acoustic noise reduction solutions Yes
*for vertical and horizontal movement

dStream Workflow (FlexStream)


FlexCoverage posterior coil Yes
FlexCoverage anterior coil (optional) Yes
FlexConnect connectors Yes
FlexTrak tabletop Yes
FlexCaddy coil storage (optional) Yes
FlexTrak patient transport system (optional) Yes
FlexTrak Mammo mammography solution (optional) Yes

SmartAssist Efficiency Assistance


SmartStart Yes
SmartSelect Yes
SmartExam (optional) Yes
SmartLine (package dependent) Yes
SmartLink (package dependent) Yes

Site Planning


Total gantry installed weight ≤ 5800 kg
Minimum siting requirement 30 m2