Respironics V60 | Patient types | | Adult | | | Pediatric | ≥ 20 kg |
| Modes | | CPAP - Continuous positive airway pressure | | S/T - Spontaneous with timed backup | | PCV - Pressure control ventilation | | AVAPS - Average volume assured pressure support (optional) | | PPV - Proportional pressure ventilation (optional) |
| Settings | | C-Flex | Off, 1 to 3 | | CPAP | 4 to 25 cm H2O | | EPAP | 4 to 25 cm H2O | | IPAP | 4 to 40 cm H2O | | I-time (inspiratory time) | 0.30 to 3.00 sec | | Max P (AVAPS maximum IPAP) | 6 to 40 cm H2O | | Min P (AVAPS minimum IPAP) | 5 to 30 cm H2O | | O2 (oxygen percent) | 21 to 100% | | Ramp Time | Off, 5 to 45 min | | Rate (respiratory rate) | 4 to 60 BPM | | Rise (rise time) | 1 to 5 | | Triggering and cycling | Auto-adaptive (Auto-Trak) or Auto-Trak Plus (optional) Trigger = normal, +1 to +7 Cycle = -2, -1, normal, +1 to +6 | | AVAPS target tidal volume | 200 to 2000 ml BTPS | | Max E | 0 to 100 cmH2O/l | | Max R | 0 to 50 cmH2O/l/s | | PPV% | 0 to 100% | | Max P (PPV maximum pressure limit) | 5 to 40 cmH2O | | Max V (PPV maximum volume limit) | 200 to 3500 ml |
| Patient Data | | Breath phase/trigger indicator | Spont, Timed, Exhale | | PIP | 0 to 50 cm H2O | | Patient/total leak | 0 to 200 l/min BTPS | | Patient Trigger | 0 to 100% | | Respiratory Rate | 0 to 90 BPM | | Ti/Ttot | 0 to 91% | | Minute volume | 0 to 99.0 l/min BTPS | | Tidal volume | 0 to 3000 ml BTPS |
| General | | Oxygen inlet pressure range | 276 to 600 kPa (40 to 87 psig) | | Weight with optional battery | 10.9 kg (24 lb) | | Weight without optional battery | 10.0 kg (22 lb) |
| Dimensions | | Height | 33.7 cm (13.3 in) | | Width | 39.4 cm (15.5 in) | | Depth | 42.9 cm (16.5 in) |
| Electrical | | AC voltage | 100 to 240 VAC | | AC frequency | 50 to 60 Hz | | AC power | 300 VA | | Battery operating time | 6 hours in normal conditions |
|
|
|